Temperature measurement
Bring every degree into focus
Sensor and transmitter options for your temperature measurement loop

Head-mounted conversion of temperature-sensor signals to 4–20 mA.
Select the measurement range, process connection and output configuration for your application.
The ST500 accepts RTD, thermocouple or linear-resistance inputs. Head mounting keeps the conversion close to the sensing point. The current output includes upper and lower overflow alarm levels, with cold-junction compensation available for thermocouple applications.
Values below describe the listed product family. Optional ranges and functions depend on the selected configuration.
| Parameter | Specification |
|---|---|
| Input signals | RTD, thermocouple, linear resistance |
| Cold-junction compensation range | −20 to 60 °C |
| Output | 4–20 mA |
| Load resistance | RL ≤ (U − 10) / 0.021 |
| Overflow alarm current | Upper 21 mA; lower 3.8 mA |
| Supply | 12–40 VDC |
| Response time | 90% of final value in 1 s |
| Operating temperature | −40 to 80 °C |
| Storage temperature | −40 to 100 °C |
| Vibration resistance | 4 g, 2–150 Hz |
Confirm the final measuring range, material compatibility, operating limits and applicable approvals for the exact configuration before ordering. Product appearance may vary by version.