Liquid analysis
A clearer view of water quality
Connect analytical measurement to treatment and process control

Panel-mounted pH and oxidation-reduction potential monitoring with isolated outputs.
Select the measurement range, process connection and output configuration for your application.
The PH6.5 displays pH or ORP together with temperature information on a monochrome LCD. Temperature compensation and isolated current and digital outputs support integration with process control equipment. The controller supports relay-based alarm functions; choose a compatible electrode and temperature element for the application.
Values below describe the listed product family. Optional ranges and functions depend on the selected configuration.
| Parameter | Specification |
|---|---|
| pH range | 0.00–14.00 pH |
| pH accuracy | ±0.02 pH |
| ORP range | −2,000 to 2,000 mV |
| ORP accuracy | ±1 mV from −1,000 to 1,000 mV; ±2 mV outside that interval |
| Input resistance | ≥10¹² Ω |
| Temperature compensation | −10 to 130 °C; NTC10K, Pt1000 or customized Pt100 |
| Analog output | 4–20 mA; maximum loop 750 Ω; ±0.2% FS |
| Digital output | Isolated RS485, Modbus-RTU |
| Operating temperature | 0–60 °C |
| Power options | 220 VAC ±10%, 110 VAC ±10% or 24 VDC ±10%; ≤6 W |
| Dimensions | 100 × 100 × 150 mm |
| Panel cutout | 92.5 × 92.5 mm |
| Protection | IP5X |
Confirm the final measuring range, material compatibility, operating limits and applicable approvals for the exact configuration before ordering. Product appearance may vary by version.