Liquid analysis
A clearer view of water quality
Connect analytical measurement to treatment and process control

Continuous monitoring of conductivity, dissolved solids and resistivity.
Select the measurement range, process connection and output configuration for your application.
The TDS210-B converts a compatible electrode signal into a local reading and process output. Electrode constant determines the usable conductivity range. The controller provides temperature compensation, isolated current output and Modbus communication, with relay outputs for control or alarm tasks.
Values below describe the listed product family. Optional ranges and functions depend on the selected configuration.
| Parameter | Specification |
|---|---|
| Variables | Conductivity, TDS, resistivity |
| Conductivity, 0.01 electrode | 0–20.00 µS/cm |
| Conductivity, 0.1 electrode | 0.20–200.0 µS/cm |
| Conductivity, 1.0 electrode | 2.00–2,000 µS/cm; maximum 20,000 µS/cm |
| Conductivity, 10.0 electrode | 0.02–20.00 mS/cm |
| Electronic-unit error | EC/TDS/ER: ±1% FS |
| Temperature compensation | −10 to 130 °C; NTC10K or Pt1000 |
| Analog output | Isolated 4–20 mA; maximum loop 750 Ω; ±0.2% FS |
| Communication | Isolated RS485, Modbus-RTU |
| Relay outputs | 2 channels; 250 VAC / 3 A |
| Power supply | 220 VAC ±10%, 50/60 Hz or 24 VDC ±10%; ≤5 W |
| Protection | IP54 |
| Dimensions | 100 × 100 × 150 mm; cutout 92.5 × 92.5 mm |
Confirm the final measuring range, material compatibility, operating limits and applicable approvals for the exact configuration before ordering. Product appearance may vary by version.